Key takeaways:

  • A new policy implemented by Stanford Health Care allowed people to use their own CGMs in the hospital.
  • Most nurses and patients preferred the new policy to mandatory finger-stick blood glucose monitoring.
Perspective from Irl B. Hirsch, MD, MACP

A change to hospital policy allowing patients to use their own continuous glucose monitors instead of finger-stick blood glucose monitoring was met favorably by a majority of patients and nurses, according to study findings.

In November 2022, Stanford Health Care put a new policy into effect allowing the use of personal CGM for inpatient care. A subcommittee of the system’s inpatient diabetes task force created a policy addressing nursing protocol for validating CGM with finger-stick blood glucose testing and the use of inpatient CGM.

Hospital corridor _42007990
A new policy from Stanford Health Care allowing patients to use their personal CGM in the hospital was well received by patients and nurses. Image: Adobe Stock

“Our patients asked if they could use their CGMs, as they were frustrated with having to do frequent finger sticks, even though they had these devices,” Michael Hughes, MD, instructor in the division of endocrinology, gerontology and metabolism at Stanford University School of Medicine, told Healio. “Some of our nurses noticed this issue too, so we sought to make it happen as a team. We found a way to reduce the number of finger sticks while still keeping an eye on the CGMs to make sure they were accurate. In the end, patients were glad to see fewer finger sticks and have their CGMs become a part of their care.”

Validating CGM measures

In a study published in Diabetes Care, Hughes and colleagues collected attempted validations of paired CGM and finger-stick blood glucose readings from the start of the policy change in November 2022 to August 2023. In order for a patient to use CGM, the device must meet the validation criteria. The criteria stated a CGM reading needed to be within 20% of a finger-stick measure if the finger-stick blood glucose was 70 mg/dL or higher or within 20 mg/dL if the finger-stick measure was less than 70 mg/dL. Mean absolute relative difference was calculated between the CGM and finger-stick measures. Accuracy was also assessed by placing measurements on a Clarke error grid. Nurses caring for patients using CGM and patients who used CGM were invited to complete a survey on the policy.

Hughes said the new protocol expands on policies implemented at other institutions by permitting the use of multiple types of CGMs.

Michael Hughes

“We also aimed to create a system that would be easy for a new intern to order, even during a late-night admission,” Hughes said. “Ultimately, this was used successfully across a fairly broad variety of hospitalizations.”

Most patients prefer CGM

There were 135 patients who used their own CGM, and 185 inpatient encounters recorded (mean age, 58 years; 43% women; 56% white). Of the patients, 51.9% had type 2 diabetes and 27.4% had type 1 diabetes. The most commonly used CGM was the Dexcom G6 (44.3% of patients) followed by the Abbott FreeStyle Libre 2 or FreeStyle Libre 14-day sensors (43.2% of patients).

Of 1,506 validation attempts, 87.8% met the criteria to allow CGM-based insulin dosing. The mean absolute relative difference between CGM and finger-stick measures was 9.6%. When validations were placed on a Clarke grid, 99.3% fell within the grid’s two most accurate zones, A and B.

Twenty-seven nurses and 46 patients completed surveys on the policy. Of the nurses, 74% said glucose management with the new policy was effective, 67% reported it was easy to use, and 63% described it as efficient. CGM was preferred over finger sticks by 16 of 20 nurse respondents.

Of the patients who completed the survey, 63% had a positive experience with the new policy, 63% said they had positive interactions with the nursing staff about CGM use and 63% said the number of diabetes management-related interruptions was reasonable. Twenty-five of 27 patients said they preferred the CGM policy over finger-stick measurements.

Moving forward

According to Hughes, the study lacked a number of data points, including the optimal timing for validation frequency and raw CGM data. However, he said the new CGM policy is feasible and could be implemented at other institutions.

“Exploring pediatrics is a very important next step, and it’s something we’re eager to pursue,” Hughes said. “The unique dynamics in pediatrics — balancing the patient’s individual autonomy with the key roles of parents and guardians — are likely to offer new insights.”