October 28, 2014
1 min read
Save

Many home blood pressure monitors may be inaccurate

You've successfully added to your alerts. You will receive an email when new content is published.

Click Here to Manage Email Alerts

We were unable to process your request. Please try again later. If you continue to have this issue please contact customerservice@slackinc.com.

Up to 15% of patients with home blood pressure monitors might be getting inaccurate readings, according to a study that will be presented at ASN Kidney Week., according to a study that will be presented at ASN Kidney Week 2014 Nov. 11–16.

Major guidelines recommend home blood pressure monitoring to guide diagnosis and treatment of patients with hypertension; however, little is known about the real-world accuracy of home blood pressure monitors that patients use.

Researchers led by Swapnil Hiremath, MD, MPH, from Ottawa Hospital and University of Ottawa, in Canada compared measurements from home blood pressure monitors with validated mercury sphygmomanometers, which are used in doctors’ offices, in 210 patients.

The investigators found that 63/210 of the home monitor systolic blood pressure readings (30%) were > 5 mm Hg different and 16 (8%) were > 10 mm Hg different from the mercury systolic blood pressure measurement. For diastolic blood pressure, these proportions were 32% (67/210) and 9% (18/210) respectively.

“Home blood pressure monitors may be inaccurate in 5% to 15% of patients, depending on the threshold for accuracy used,” said. Hiremath. “We recommend all patients with home monitors get them validated with their health care providers at least once.”

Study: “Are Home Blood Pressure Monitors Accurate Compared to Validated Devices?” (Abstract SA-PO187)